Performance of devices in medical applications (in vitro)
We determine the following performance parameters e.g. of hemodialyzers:
- Clearance of low molecular weight solutes in aqueous solutions according to EN1283
- Clearance of low molecular weight plasma proteins (LMWP) from blood and plasma
- Determination of sieving coefficients of plasma proteins from plasma or human blood according to EN1283
- Albumin loss into dialysate during hemodialysis und hemodiafiltration
- Determination of transmembrane pressure (TMP) in aqueous solution, plasma and human blood.
- Determination of ultrafiltration rate (UFR) and ultrafiltration coefficient (KUF) in aqueous solution, plasma and human blood.
LMWP clearances or albumin loss is determined using freshly donated human blood or plasma supplemented with concentrated hemofiltrate obtained from end-stage renal disease (ESRD) patients. By this approach donor blood gets closer to the properties of uremic blood (e.g. with regard to the concentration of uremic toxins like β2-microglobulin).
We also offer standardized methods for clearance of plasma proteins, such as:
- ß2-Microglobulin
- Myoglobin
- Cystatin C
- Retinol binding protein (RbP)
- α1-Microglobulin
- etc. (directory of services)
in human donor blood, plasma or albumin solution. In parallel to the clearance measurements in human blood, plasma or albumin solution we can determine albumin loss into dialysate in vitro.
Determination of sieving coefficients is carried out in in vitro-experiments using hemodialyzers, hemodiafilters, hemofilters or hemoconcentrators and a dialysis monitor in either human donor blood, plasma or albumin solution.
We determine sieving coefficients of e.g. the following proteins:
- ß2-Microglobulin
- Myoglobin
- Cystatin C
- Retinol binding Protein (RbP)
- α1-Microglobulin
- Albumin
- other
in plasma or human blood according to EN 1283 or ANSI/AAMI RD16.
We are also experienced in characterizing highly porous membranes compared to dialysis membranes. Highly porous membranes are used in high cut-off filters, in artificial liver or in plasma fractionation and plasma separation. In standard experiments the following proteins, resp. particles are determined:
- Albumin
- IgG
- HDL
- Fibrinogen
- LDL
- IgM
TMP is the primary factor determining (ultra)filtration rate. During filtration, the feed side of the membrane is showing higher pressure than the permeate side. This pressure difference forces fluids such as plasma water to pass the membrane. Determination of transmembrane pressure is carried out in in vitro-experiments but also during clinical studies using a custom designed four point pressure measurement (P bloodin, P bloodout, P dialysatein, P dialysateout) data acquisition system.
Ultrafiltration rate is defined as the hourly (ultra)filtration in mL, per m² dialyzer area and mm Hg TMP. The efficacy of a particular ultrafiltration device in producing ultrafiltrate is called ultrafiltration coefficient (KUF). It is expressed as the ultrafiltration per hour in mL at a TMP of one mm Hg. Determination of UFR and KUF is carried out in in vitro-experiments and clinical studies and based on our four point TMP measurement. For protein containing solutions oncotic pressure can be taken into account by using the formula of Landis and Pappenheimer.